JWI-501
3-9-1-2
When choose on and press <TARE>, now enter to set limit. “under” lamp is
lighting , max figure is blinking. Press<SET/0~9> once to add 1. Press <TARE> to
confirm. When finishing, auto exit and set “over”(same as set under limit). After finish, turn
back to normal weighing mode.
3-9-1-3 When low limit > high limit, it displays off, then exit.
3-9-1-4 Alarming indicate: m is the weight
When m≥high limit, “over” lamp is lighting and flickering.
When low limit <m<high limit, “accept” lamp is lighting (not flickering).
When m≤low limit, “under” lamp is lighting.
Note:If the stable indication does not appear, no lamp lights.
3-9-2
UnItS : choose unit. Default = kg
3-9-2-1 When it displays units, press <TARE> to enter. Press <SET/0~9> to change.
Press <TARE> to confirm. Units:h9(kg), pcs, (counting), lb.lb(decimal pounds)
3-9-2-2 Counting:When it displays “pcs”, press<TARE>. Now “pcs” lamp lighting.
Press and hold <ON/ZERO>, it displays count and return to 0. Then put the sample (as
much as possible, but don’t more than max.capacity). After stable, press<TARE>. Press
<SET/0~9> to add 1 and press<TARE>to choose. After finishing, it displays pcs and
“pcs” lamp lighting.
exit counting: choose a weighing unit.
Note: Quantities of max less 30000
3-9-3 A-OFF: choose auto power off. Default = n
When it displays A-OFF, press <TARE> to enter. Press <SET/0~9> to choose “n” or “y”,
n = not in auto power off situation; y = after 10 minutes in zeroing, auto-off.
3-9-4
FILt : set display mode. Default = 3
When it displays FILt, press <TARE>to enter. Press <SET/0~9> to choose filt-1 or filt-2
or filt-3 or filt-4. Press <TARE>to confirm.
filt-1:When add weight to scale, it will show the value till stable.
filt-2:When add weight to scale, it will show slow.
filt-3:When add weight to scale, it will show fast. (Recommend)
filt-4:When add weight to scale, it will show much faster.