EasyManua.ls Logo

Jain Technology XONIC 100L - Total Flow Set; Alarm Settings; Calibration

Jain Technology XONIC 100L
42 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Xonic 100L, 2021
18
Total Flow Set If user need to change flow total, can change by keypad.
Alarm Xonic 100 has alarm functions. User can set HIGH FLOW, LOW FLOW
alarms.
Calibration In case the user has a calibration instrument or other laboratory instrument to
test the Xonic 100, then they can use this menu to calibrate for best accuracy,
user can select calibration menu.
OPERATE
1 . UPPER FLOW LIMIT
2 . LOWER FLOW LIMIT
3 . DEAD ZONE
4 . FLOW AVERAGE TIME
5 . TOTAL FLOW SET
6 . ALARM
7 . CALIBRATION
8 . ENABLE AGC
9 . DAMPING
0 . FIX RISC.
FLOW AVERAGE TIME
sec
10
OPERATE
1 . UPPER FLOW LIMIT
2 . LOWER FLOW LIMIT
3 . DEAD ZONE
4 . FLOW AVERAGE TIME
5 . TOTAL FLOW SET
6 . ALARM
7 . CALIBRATION
8 . ENABLE AGC
9 . DAMPING 1 . POSITIVE SET
0 . FIX RISC. 2 . NEGATIVE SET
TOTAL FLOW SET
OPERATE
1 . UPPER FLOW LIMIT
2 . LOWER FLOW LIMIT
3 . DEAD ZONE
4 . FLOW AVERAGE TIME
5 . TOTAL FLOW SET
6 . ALARM
7 . CALIBRATION
8 . ENABLE AGC
9 . DAMPING 1 . LOW FLOW
0 . FIX RISC. 2 . HIGH FLOW
ALARM
OPERATE
1 . UPPER FLOW LIMIT
2 . LOWER FLOW LIMIT
3 . DEAD ZONE
4 . FLOW AVERAGE TIME
5 . TOTAL FLOW SET
6 . ALARM METHOD
7 . CALIBRATION 1 . NO CALIBRATION
8 . ENABLE AGC 1 . METHOD 2 . MULTI-POINTS
9 . DAMPING 2 . MULTI-POINT SET 3 . Kc CALIBRATION
0 . FIX RISC. 3 . Kc SET 4 . Kc and MULTI
CALIBRATION