JEOL 1010 - TEM Alignment and Diaphragm Adjustment; Condenser Diaphragm Settings

Default Icon
15 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Version 2017-01-19 for GI use
2. Remove the upper and the lower condenser diaphragms by switching the level
c to the right
3. Go up and down with brightness (12). Adjust over and underfocus. If the A well
centered filament gives a centered beam position swings, adjust the centering with the
screws shift 10 and shift 13