Julabo F32-HE - 6.2.2. Switching on the Cooling Machine

Julabo F32-HE
88 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating procedures
34
6.2.2. Switching on the Cooling Machine
Switching on:
x Switch on the cooling machine using the switch (1b) .
L Control of the cooling machine:
With the mains switch (1b) turned on, the circulator automatically
switches the cooling machine off and on.
x It is switched off, if:
- the actual working temperature is increased by >30 °C (cooling is
not required).
- the heater operates at full power (>800 W) for longer than 5
minutes.
x It is switched on, if:
- cooling is necessary for maintaining the bath temperature.
After switch-off, the cooling machine automatically switches on only
after a delay of 5 minutes for protecting the cooling compressor.
L To save energy, turn off the cooling machine with the mains switch
(1b) whenever cooling is not required.