EasyManua.ls Logo

Julabo F32-HE - Page 67

Julabo F32-HE
88 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
HE
67
BAND HIGH/LOW – Band limitation
The band limitation is active during external control. Varied, practice-
oriented setting are feasible for heat-up and cool-down phases.
Setting range: 0 °C ... 200 °C
BAND HIGH and BAND LOW allow for the limitation of the difference
between the temperatures in the internal bath and the external system to
any maximum value for the heat-up and cool-down phase.
During the heat-up phase this difference value is always added to the
actual external temperature. During the cool-down phase the difference
value is subtracted.
25
50
10 20 30 40 50 60 10 20 30 40 50 60 10
[min]
75
100
150
2
00
°C
t
external system
Internal bath
BAND HIGH
BAND LOW
Sense of a band limitation:
; Protection of objects and samples
by gentle temperature control
; Protection of e.g. glass reactors
from thermal shock.
www.calcert.com sales@calcert.com1.888.610.7664