6.3.3. Setting the safety temperatures
F G
)
This safety function is independent of the control circuit.
n
Press the desired setpoint key (F, G).
Follow the instructions under section 6.3.1.
page 24
o
p
q
Recommendation:
Set the high temperature limit at least 5 K above the actual
bath temperature.
Set the low temperature limit at least 5 K below the setpoint.
+
ACTUAL
SET
When the temperature of the bath liquid reaches the limits of
the safety values, a complete shutdown of the circulating
pump, heater and cooling compressor is effected.
The alarm light illuminates and an audible signal is
triggered. An error message appears on the MULTI-
DISPLAY (LED 2)
(see page 29).
Turn the mains switch (1.) off and on. The alarm state is
cancelled and the circulator is put back into operation.
(Switching on - see page 22).
6.3.4. Low liquid level protection
+
ACTUAL
SET
As soon as the bath liquid falls below the "MIN" level
(10.), a complete shutdown of the circulating pump,
heater, and cooling compressor is effected.
The alarm light illuminates and an audible signal is
triggered. An error message appears on the MULTI-
DISPLAY (LED 2)
(see page 29).