Operator responsibility – Safety instructions
8
2.4. Technical specifications
FC(W)600 FC(W)600S
Working temperature range °C -20 ... 80 -10 ... 80
Cooling capacity °C 20 5 -10 20 5 -10
(water-glycol) kW 0.6 0.4 0.21 0.5 0.3 0.1
Refrigerant R134a
Heater capacity kW 1.2
Pump capacity:
Pressure max. psi 7.25 17.4
Flow rate max. Lpm 20 / 14 22 / 15
with tubing connections mm dia 13,5/9,5 13,5/9,5
Pump connections M16x1 M16x1
Noise level, 1 m distance dBA 51 54
Filling volume l 6 ... 8
Dimensions (WxLxH) mm 350 x 540 x 490
Ambient temperature °C 5 ... 40
Shipping weight kg 48 52
Mains power connection V/Hz 230/50 or 208 - 230 / 60
Total power consumption W 2000 2300
FC(W)1200 FC(W)1200S
Working temperature range °C -20 ... 80 -15 ... 80
Cooling capacity °C 20 5 -10 20 5 -10
(water-glycol) kW 1.3 0.75 0.37 1.2 0.65 0.26
Refrigerant R134a
Heater capacity kW 1.2
Pump capacity:
Pressure max. bar 0.5 1.2
Flow rate max. l/min 20 / 14 22 / 15
with tubing connections mm 13,5/9,5 13,5/9,5
Pump connections M16x1 M16x1
Noise level, 1 m distance dBA 53 57
Filling volume l 8 ... 11
Dimensions (WxLxH) mm 460 x 610 x 490
Ambient temperature °C 5 ... 40
Shipping weight kg 60 66
Mains power connection V/Hz 230/50 or 208 - 230 / 60
Total power consumption W 2300 2800
All data have been determined at mains voltage: 230 V / 50 Hz
ambient temperature: 20 °C bath liquid: water-glycol