EMC performance
WARNING
• This product requires special EMC precautions and needs to be installed and used
according to the following information.
• Use of accessories, transducers and cables other than those specied or provided by
the manufacturer of this equipment could result in increased electromagnetic
emissions or decreased electromagnetic immunity of this equipment and result in
improper operation.
• Cable length: 3.0 m or less.
• Use of this equipment adjacent to or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, this equipment
and the other equipment should be observed to verify that they are operating
normally.
• Portable RF communications equipment (including peripherals such as antenna cables
and external antennas) should be used no closer than 30 cm (12 inches) to any part of
the model CL-S500 including cables specied by the manufacturer. Otherwise,
degradation of the performance of this equipment could result.
• Any additional equipment connected to this product needs to comply with the
requirements of IEC 60601-1-2:2014/EN 60601-1-2:2015.
Environments of Intended Use
The model CL-S500 is intended to be used in Home healthcare environment such as
residences and nursing homes.
Guidance and manufacturer’s declaration -
Electromagnetic compatibility
Guidance and manufacturer’s declaration – electromagnetic emissions
The model CL-S500 is intended for use in the electromagnetic environment specied below.
The customer or the user of the model CL-S500 should assure that it is used in such an
environment.
Emissions test Compliance
Electromagnetic environment
– guidance
RF emissions
CISPR 11/EN 55011
Group 1 The model CL-S500 uses RF energy only
for its internal function. Therefore, its
RF emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11/EN 55011
Class B The model CL-S500 is suitable for use in
all establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class D
Voltage uctuations/icker
emissions
IEC 61000-3-3
Complies