Guidance and manufacturer’s declaration – electromagnetic immunity
The model CL-S50 0 is intended for use in the electromagnetic immunity environment (Home
healthcare environment dened in IEC 60601-1-2) specied below.
The customer or the user of the model CL-S50 0 should assure that it is used in such an
environment.
Immunity IEC 60601 test level
Compliance
level
Electromagnetic environment
– guidance
Proximity
electromagnetic eld
from RF wireless
communication
equipment
(specied the table
below)
Portable and mobile RF
communications equipment
should be used no closer to any
part of the model CL-S50 0,
including cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Conducted RF
IEC 61000-4-6
3 V
0.15 MHz to 80 MHz
6 V
ISM bands and
amateur radio
between
150 kHz and 80 MHz
80 % AM at 1 kHz
3 V
0.15 MHz to 80 MHz
6 V
ISM bands and
amateur radio
between
150 kHz and 80 MHz
80 % AM at 1 kHz
Recommended separation
distance
d
= 1.2
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to 2.7 GHz
80 % AM at 1 kHz
10 V/m
80 MHz to 2.7 GHz
80 % AM at 1 kHz
d
= 1.2 80 MHz to 800 MHz
d
= 2.3 800 MHz to 2.7 GHz
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and
d
is
the recommended separation
distance in meters (m).
Field strengths from xed RF
transmitters, as determined by an
electromagnetic site survey,
a
should be less than the
compliance level in each
frequency range.
b
Interference may occur in the
vicinity of equipment marked with
the following symbol: