WATER TRAP MAINTANENCE
1)
Carefully remove
water trap from housing
2)
Pull reservoir vertically
from filter holder
3)
Rotate top part of
filter housing 30°
anti-clockwise
4)
Pull particulate filter
receiver vertically from
water stop filter receiver
KANE988 Water Stop Technology
Your analyser has a water trap & particle filter to stop appliance flue gas
water vapour & dust entering your analyser.
However, some boilers produce much higher volumes of water vapour which
can affect your analyser.
Your analyser has a water stop filter with hydrophobic technology located
inside the water trap to stop water vapour.
The water stop filter sits in a filter carrier located above the particle
filter inside the water trap.
You must replace water stop & particle filters when wet, dirty or your analyser
displays LOW FLOW.
To replace:
KANE988 MANUAL 35