Limitations
The product is not suitable for use in the environments with a high concentration of oxygen and/or
flammable liquids and/or flammable gas; do not use with equipment for electrosurgery or shortwave or
microwave therapy; the device may be interfered by other equipment, even if that other equipment
complies with CISPR EMISSION requirements.
Emissions
Voltage Fluctuation and Flicker Emissions
Immunity
IEC 60601-1-2
Test Levels
Contact: ± 8kV
Air: ± 2kV, ± 4kV, ± 8kV, and ± 15kV
No performance
degradation observed.
80MHz to 2.7GHz, 10V/m; @ 80%,
1kHz and Table 9 of Clause 8.10
No performance
degradation observed.
Electrical Fast
Transients/Bursts
AC Mains: 2kV, 100kHz PRF
DC Mains: NA
No performance
degradation observed.
AC Mains: ± 0,5 kV and ± 1 kV Line to
Line, ± 0,5 kV, ± 1 kV and ± 2 kV Line
to Ground
DC Mains: NA
IO Lines: NA
No performance
degradation observed.
AC Mains: 150kHz to 80MHz, 3 Vrms
outside ISM and amateur radio bands,
6 Vrms in ISM and amateur radio
bands: @ 80%, 1kHz
IO Lines: 150kHz to 80MHz, 3 Vrms
outside ISM and amateur radio bands,
No performance
degradation observed.