7 Technical data
36 FOCUS
Table 1: Electromagnetic emissions IEC 60601-1-2 Ed4
The device is suitable for use in the specified electromagnetic environment. The purchaser or user
of the device should assure that it is used in an electromagnetic environment as described below:
Emissions Test Compliance Electromagnetic Environment
Radio-Frequency
Emissions CISPR11
Group 1 The device uses RF energy only for its internal func-
tion. Therefore, the RF emission is very low and not
likely to cause any interference in nearby electronic
equipment.
Radio-Frequency
Emissions CISPR11
Class B
The device is suitable for use in both at professional
healthcare (hospitals/large clinics) facility
environment and home healthcare (clinics in
domestic establishments and those directly
connected to the public low-voltage power supply)
environment.
Exceptions for professional healthcare facility
environment :
Not to be used or installed near active HF SURGICAL
EQUIPMENT and the RF shielded room of an ME
SYSTEM for magnetic resonance imaging, where the
intensity of EM DISTURBANCES is high.
Harmonic emissions
IEC61000-3-2
IEC61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC61000-3-3
Complies