7 Technical data
FOCUS 41
INSTALLATION REQUIREMENTS & ENVIRONMENT
CONTROL:
In order to minimize interference risks, the following
requirements shall apply.
Cables shielding & grounding
All interconnect cables to peripheral devices must meet the
requirements given in Technical specifications. Use of
incorrect cables may result in the device causing
radio frequency interference.
Electrostatic discharges environment &
recommendations
In order to reduce electrostatic discharge interference, a
charge dissipative floor should be installed to prevent
charge accumulation.
• The dissipative floor material must be connected to
the system reference ground, if applicable.
• Relative humidity must be maintained above 30
percent.
Stacked components & equipment
WARNING! The device should not be used adjacent to or
stacked with other equipment; if adjacent or stacked use is
necessary, the device should be observed to verify normal
operation in the configuration in which it will be used.
Interference may occur in the vicinity of equipment marked
with the following symbol: