EasyManuals Logo

Keithley 2001 Quick Reference Guide

Keithley 2001
94 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #13 background imageLoading...
Page #13 background image
10
+000.0000 mVDC
Rdg#+00000 @Time =+000.000000 sec
2. Use the RANGE keys to increment and decrement
the reading number and scroll through the buffer.
3. The NEXT and PREV DISPLAY keys can be used
while recalling readings to access additional buffer
data, such as maximum and minimum readings.
The following code fragment configures the Model
2001 for high speed DC voltage readings, stores 100
readings, and sends the readings over the bus:
100 OUTPUT 716;“:syst:pres”
110 OUTPUT 716;“:volt:dc:nplc 0.01”
120 OUTPUT 716;“:trac:clear”
130 OUTPUT 716;“:trac:feed calc”
140 OUTPUT 716;“:trac:egr full”
150 OUTPUT 716;“:trac:poin 100”
160 OUTPUT 716;“:trac:feed:cont next”
170 OUTPUT 716;“:trac:data?”
180 ENTER 716;A$
190 PRINT A$

Other manuals for Keithley 2001

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2001 and is the answer not in the manual?

Keithley 2001 Specifications

General IconGeneral
Model2001
CategoryMultimeter
ManufacturerKeithley
TypeDigital Multimeter
Digits of Resolution7.5
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
ConnectivityGPIB, RS-232
DisplayVFD
Resistance Accuracy0.008% + 2.0 mΩ
Frequency Accuracy±0.01%
DC Current Range10 μA to 3 A
AC Current Range10 μA to 3 A
Power Supply100 V to 240 V AC, 45 Hz to 66 Hz