Model 6485 and 6487 User’s Manual Range, Units, Digits, Rate, and Filters 4-5
Median filter
The median filter is used to determine the "middle-most" reading from a group of readings
that are arranged according to size. For example, assume the following readings:
20mA, 1mA, 3mA
The readings are rearranged in an
ascending order as follows:
1mA, 3mA, 20mA
From the above readings, it is apparent that
3mA is the median (middle-most) reading.
The number of sample readings used for the median calculation is determined by the
selected rank (1 to 5) as follows:
Sample readings = (2 × R) +
1
where; R is the selected rank (1 to 5)
Note that the median filter throws out the oldest of
the ranked values. It is especially use-
ful for noise characterized by
unwanted spikes.
Digital filter
Digital filter types
An additional filter parameter is type (type is either moving or repeating).
Moving Filter
Every time a reading conversion occurs, the readings in the stack are
averaged to yield a single filtered reading. The stack type is first-in, first-out. After the
stack fills, the newest reading conversion replaces the oldest. Note that the instrument
does not wait for the stack to fill before releasing readings.
Repeating Filter
Takes a selected number of reading conversions, averages them, and
yields a reading. It then flushes its stack and starts over.

Table of Contents

Other manuals for Keithley 6487

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6487 and is the answer not in the manual?

Keithley 6487 Specifications

General IconGeneral
TypePicoammeter/Voltage Source
Current Measurement Range20 fA to 20 mA
InterfacesGPIB, RS-232
Voltage Burden< 200 µV
Input Impedance>10^14 Ω
Power100 V / 120 V / 220 V / 240 V
Basic Current Measurement Accuracy0.4%

Related product manuals