6-6 Buffer and Sweeps Model 6485 and 6487 User’s Manual
Programming example
The following command sequence performs a sweep from 1V to 10V in 1V increments:
*RST ' Return 6487 to RST defaults.
SOUR:VOLT:SWE:STAR 1 ' Start voltage = 1V.
SOUR:VOLT:SWE:STOP 10 ' Stop voltage = 10V.
SOUR:VOLT:SWE:STEP 1 ' Step voltage = 1V.
SOUR:VOLT:SWE:DEL 0.1 ' 0.1s delay.
ARM:COUN INF ' Reset arm count.
FORM:ELEM READ,VSO ' Select reading, voltage source data.
SOUR:VOLT:SWE:INIT ' Arm sweep, put source in operate.
SYST:ZCH OFF ' Turn off zero check.
INIT ' Trigger sweep.
TRAC:DATA? ' Request all stored readings.

Table of Contents

Other manuals for Keithley 6487

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6487 and is the answer not in the manual?

Keithley 6487 Specifications

General IconGeneral
TypePicoammeter/Voltage Source
Current Measurement Range20 fA to 20 mA
InterfacesGPIB, RS-232
Voltage Burden< 200 µV
Input Impedance>10^14 Ω
Power100 V / 120 V / 220 V / 240 V
Basic Current Measurement Accuracy0.4%