Model 6485 and 6487 User’s Manual Relative, mX+b, m/X+b (reciprocal), and log 5-3
Programming example — relative
This program fragment establishes a 1μA baseline for measurements:
CALC2:NULL:OFFS 1e-6 ' Set Rel value of 1μA.
CALC2:NULL:STAT ON ' Enable Rel.
SYST:ZCH OFF ' Turn off zero check.
INIT ' Trigger reading.
CALC2:DATA? ' Request Rel’ed reading.
mX+b, m/X+b (reciprocal), and Logarithmic
mX+b and m/X+b
The following math operations manipulate normal display readings (X) mathematically
according to the following calculations:
Y = mX+b
Y = m/X+b
where: X is the normal displ
ay reading.
m and b are user-entered constants
for scale factor and offset.
Y is the displayed result.
NOTE Cha
nging the “m” or “b” for mX+b also changes it for m/X+b.
Logarithmic
This calculation converts input readings to logarithm base 10 values. The calculation is
performed as follows:
10
Xlog Y=
where: X is the input reading.
Y is the logarithmic result.
For example: Assume that exactly 1mA is being measured by the Model 6485/6487.
1.000000mA
10
log 3=
NOTE This calculation uses the absolute value of the normal input reading as the log
of a negative number cannot be computed.

Table of Contents

Other manuals for Keithley 6487

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6487 and is the answer not in the manual?

Keithley 6487 Specifications

General IconGeneral
TypePicoammeter/Voltage Source
Current Measurement Range20 fA to 20 mA
InterfacesGPIB, RS-232
Voltage Burden< 200 µV
Input Impedance>10^14 Ω
Power100 V / 120 V / 220 V / 240 V
Basic Current Measurement Accuracy0.4%

Related product manuals