6: Leakage current and insulation resistance Model 2450 Interactive SourceMeter® Instrument
User's Manual
6-8 2450-900-01 Rev. C / December 2013
Figure 28: Leakage current test results
Measuring insulation resistance
The following applications demonstrate how to use the Model 2450 to measure insulator resistance.
The applications shows how to use the front-panel interface, the remote interface using the SCPI
command set, and the remote interface using the Test Script Processor (TSP
®
) command set.
You make insulation resistance measurements between traces on a printed circuit board and
between conductors in cables and connectors.
This application sets the Model 2450 to source 20 V and makes 10 resistance readings with a 100 ms
interval time. Once the measurements are made, the output is turned off.
For this test, you will:
Reset the instrument
Set the instrument to read the rear terminals
Select the source voltage function and measure resistance function
Set the magnitude of the voltage source output
Turn on autoranging
Use the Simple Loop trigger model template to set the numbers of readings and interval time
Turn on the source output
Make readings
Turn off the source output