4-2 Controlling Multiple SourceMeters Series 2600 System SourceMeters User’s Manual
2600S-900-01 Rev. A / May 2006 Return to In this section:
How do I set up the TSP-Link system?
Up to 64 TSP-Link-enabled instruments (e.g., Series 2600 SourceMeters) can be
connected together to form a TSP-Link system.
Step 1: Connect the TSP-Link system
Reference Refer to “Connections” in Section 9 of the Series 2600 Reference
Manual for more information on TSP-Link connections.
A connection example for an expanded system is shown in Figure 4-1. All the
TSP-Link-enabled instruments in the system are daisy-chained together using
LAN crossover cables. In this example, the unit connected to the PC is the Master
and the other units are its Slaves. Note that the Master does not have to be
assigned as Node 1, nor does it have to be at the end of the node chain.
Figure 4-1
TSP-Link connections
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
IEEE-488
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
LINE RATING
100-240VAC
50, 60Hz
240VA MAX.
DIGITAL I/O
RS-232
TSPlink
LINE FUSE
SLOWBLOW
3.15A, 250V
CAT I
SS
LO LO HIHI GGGG
CHANNEL A
MADE IN
U.S.A.
!
!
!
CAT I
S
S
LO
LO
HI
HI
G
G
G
G
CHANNEL B
!
U
L
C
US
LISTED
SourceMeter
4ZA4
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
IEEE-488
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
LINE RATING
100-240VAC
50, 60Hz
240VA MAX.
DIGITAL I/O
RS-232
TSPlink
LINE FUSE
SLOWBLOW
3.15A, 250V
CAT I
SS
LO LO HIHI GGGG
CHANNEL A
MADE IN
U.S.A.
!
!
!
CAT I
S
S
LO
LO
HI
HI
G
G
G
G
CHANNEL B
!
U
L
C
US
LISTED
SourceMeter
4ZA4
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
IEEE-488
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
LINE RATING
100-240VAC
50, 60Hz
240VA MAX.
DIGITAL I/O
RS-232
TSPlink
LINE FUSE
SLOWBLOW
3.15A, 250V
CAT I
SS
LO LO HIHI GGGG
CHANNEL A
MADE IN
U.S.A.
!
!
!
CAT I
S
S
LO
LO
HI
HI
G
G
G
G
CHANNEL B
!
U
L
C
US
LISTED
SourceMeter
4ZA4
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
IEEE-488
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
LINE RATING
100-240VAC
50, 60Hz
240VA MAX.
DIGITAL I/O
RS-232
TSPlink
LINE FUSE
SLOWBLOW
3.15A, 250V
CAT I
SS
LO LO HIHI GGGG
CHANNEL A
MADE IN
U.S.A.
!
!
!
CAT I
S
S
LO
LO
HI
HI
G
G
G
G
CHANNEL B
!
U
L
C
US
LISTED
SourceMeter
4ZA4
Node 1 Node 2 Node 3 Node 64
Host
PC
RS-232
GPIB
or
LAN crossover cables
Type: Category 5e or higher.
Length: 3 meters maximum between nodes.
TSP-Link connectors
(2 per instrument)
NOTE The PC is not needed for
stand-alone systems.
Step 2: Assign node numbers
Reference Refer to “Initialization” in Section 9 of the Series 2600 Reference
Manual for more information on assigning node numbers.
Each instrument in the TSP-Link system must have a unique node number
assigned to it. The node number for each instrument is stored in its non-volatile
memory and will not be lost when the instrument is turned off.
The TSP-Link node number is set from the MAIN MENU of each instrument and is
summarized in
Table 4-1.

Table of Contents

Other manuals for Keithley SourceMeter 2611

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2611 and is the answer not in the manual?

Keithley SourceMeter 2611 Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2611
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals