6.7.2 Models CPB-DM (verified models)
The adjustment is locked for verified balances. Carrying out adjustment
requires that the seal is destroyed and the adjusting switch is pressed when
turning on the scale. For position of adjusting switch, see chap. 6.9.1.
Attention:
After destruction of the seal the balance must be re-verified by an
authorised agency and a new verification wire/seal mark fitted before it can
be reused for applications subject to verification.
Operation Indication
Switch-on balance and during the selftest
actuate the adjustment switch and press
the TARE button
Use the number keys to enter password:
either
Default password "0000"
or
Personal password, enter under
function [F6 PKn ] see chap.12
Confirm with TARE button
CPB-N / CPB-DM-BA-e-1424 17