EasyManua.ls Logo

KERN EG 620-3NM - Parameters for the Serial Interface

KERN EG 620-3NM
57 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
EW/EG-N/EWB-e-1627
50
7.2.2 Parameters for the serial interface
(excl. EWB models)
Function
Display
Choice
Description of the choice pos-
sibilities
Output format at inter-
face
7
I.F.
0
Interface not active
1
6-digit data format
2
7-digit data format
Output condition at
interface
71.
o.c. 0 No data issued.
(Menu setting “7 I.F. [1] or
[2]“ only )
1 Continuous serial output.
2 Continuous serial output upon
stabilised display.
3 Output following printing of
PRINT/M.
4 Automatic output upon stable
weighing value. The first value to
stabilise is adopted if this is -0.00
or
less. No new output until weight
is removed and a new load ap-
plied.
5 Output upon stabilisation, no out-
put if data unstable.
6 Output upon stabilisation, con-
stant output if data unstable.
7 Output following printing of
PRINT/M.
Baud rate
72.
b.L.
1
1200 bps
2
2400 bps
3
4800 bps
4
9600 bps
Parity
73.
PA.
0
No parity bit
(Menu setting
1
Uneven parity
“7 I.F. 2“ only )
2
Even parity

Table of Contents