EasyManuals Logo

KERN EMS User Manual

KERN EMS
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #36 background imageLoading...
Page #36 background image
10 EMS-BA-e-1213
If the batteries are run down, "LO" appears in the display. Press ON/OFF-key and
replace the batteries immediately.
If the balance is not used for a longer time, take out the batteries and store them
separately. Leaking battery liquid could damage the balance.
If there exists an optional rechargeable battery, it has to be connected in the battery
compartment via a separate plug-in socket. Now the mains adapter delivered with
the rechargeable battery must be applied.
5.5 Initial Commissioning
In order to obtain exact results with the electronic balances, your balance must have
reached the operating temperature (see warming up time chap. 1). During this
warming up time the balance must be connected to the power supply (mains,
accumulator or battery).
The accuracy of the balance depends on the local acceleration of gravity.
Strictly observe hints in chapter Adjustment.
5.6 Adjustment
As the acceleration value due to gravity is not the same at every location on earth,
each balance must be coordinated - in compliance with the underlying physical
weighing principle - to the existing acceleration due to gravity at its place of location
(only if the balance has not already been adjusted to the location in the factory). This
adjustment process must be carried out for the first commissioning, after each
change of location as well as in case of fluctuating environment temperature. To
receive accurate measuring values it is also recommended to adjust the balance
periodically in weighing operation.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN EMS and is the answer not in the manual?

KERN EMS Specifications

General IconGeneral
CalibrationExternal
DisplayLCD
MaterialStainless steel
Weighing unitsg, oz
Power supplyMains adapter
Plate size150 mm
Permissible ambient temperature10°C to 40°C

Related product manuals