EMS-BA-e-1213 11
5.7 Adjustment
The adjustment should be made with the recommended adjustment weight (see
chap. 1 "Technical data"). Adjustment is also possible with the weights of other
nominal values (see table 1), but not the optimum for measuring technique.
Procedure when adjusting:
Observe stable environmental conditions. A warming up time (see chapter 1) is
required for stabilization.
Turn on balance by pressing the ON/OFF key.
Press the MODE key and keep it pressed, in the display appears shortly
CAL“. After that the exact size appears flashing in the display of the selected
adjustment weight.
Now set the adjusting weight in the centre of the weighing plate.
Press the SET key. Short time later there appears „CAL F“, then the automatic
return to the weighing mode. In the display there appears the value of the
adjustment weight.
An error during adjustment or the use of an incorrect adjusting weight will
result in an error message „CAL E“. Repeat adjustment.
Keep the adjustment close to the balance. Daily control of the weighing exactness is
recommended for quality-relevant applications.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN EMS and is the answer not in the manual?

KERN EMS Specifications

General IconGeneral
CalibrationExternal
DisplayLCD
MaterialStainless steel
Weighing unitsg, oz
Power supplyMains adapter
Plate size150 mm
Permissible ambient temperature10°C to 40°C

Related product manuals