- 21 - KFB-TM/KFN-TM-SH-e-1410
Stable weighing value for the weight is sent via
the RS232 interface
Weighing value for the weight (stable or
unstable) is sent via the RS232 interface
US,GS 1.342KG
ST,GS 1.000KG
No data are sent, the balance carries out the
tare function.
No data are sent, the zero-display appears.
Quantity will be sent via the RS232-interface
8 Instant help
In case of an error in the program process, briefly turn off the display unit and
disconnect from power supply. The weighing process must then be restarted from the
beginning.
Help:
The displayed weight does not
glow.
The display unit is not switched on.
Mains power supply interrupted (mains cable defective).
Power supply interrupted.
(Rechargeable) batteries are inserted incorrectly or
empty
No (rechargeable) batteries inserted.
The displayed weight is
permanently changing
Weighing pan has contact with other objects.
Electromagnetic fields / static charging (choose different
location/switch off interfering device if possible)
The weighing result is
obviously incorrect
The display of the balance is not at zero
Adjustment is no longer correct.