EasyManuals Logo

Keysight Technologies 34980A User Manual

Keysight Technologies 34980A
149 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #53 background imageLoading...
Page #53 background image
Calibration Procedures 3
Keysight 34980A Service Guide 53
Gain Adjustment Procedure
Adjustment for each function should be performed only in the order shown in the
performance verification table. The performance verification tables used for gain
adjustments start on page 44.
Review the “DMM Test Considerations” on page 42 and “Gain Adjustment
Considerations” on page 51 sections before beginning this test.
Configuration: DC functions — 6½ digits
AC functions — LF 3 HZ:SLOW
1 Configure each function and range shown in the gain verification tables
(starting on page 44).
2 Apply the input signal shown in the “Input” column of the appropriate
verification table.
3 Send the actual input value to the instrument using the CALibration:VALue
<value> command.
4 Calibrate the instrument using the CALibration? command.
5 Perform the appropriate Gain Verification Test to check the calibration results.
6 Repeat steps 1 through 5 for each gain verification test point shown in the
tables.
NOTE
Always complete tests in the same order as shown in the appropriate verification
table.
NOTE
Each range in the gain adjustment procedure takes less than 20 seconds to
complete.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies 34980A and is the answer not in the manual?

Keysight Technologies 34980A Specifications

General IconGeneral
BrandKeysight Technologies
Model34980A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals