Calibration Procedures 3
Keysight 34980A Service Guide 55
Plug-in Modules
Plug-in Module Test Considerations
For optimum performance, all test procedures should comply with the following
recommendations:
Assure that the calibration ambient temperature is stable and between 18 °C
and 28 °C. Ideally the calibration should be performed at 23 °C ± 1 °C.
Assure ambient relative humidity is less than 80%.
Install the plug-in module and allow a 1 hour warm-up period before
verification or adjustment.
Use shielded twisted pair PTFE insulated cables to reduce settling and noise
errors. Keep the input cables as short as possible.
Remove all user wiring and connections from the plug-in modules before
verification or adjustment.
Use 4-wire Ohms measurement techniques for checking relay contact
resistance. Check directly at the terminals where possible.