Calibration Procedures 3
Keysight 34980A Service Guide 59
.
Output Voltage Quick Check Error from Nominal (90 day)
16 V Q ± 11 mV
12 V ± 9 mV
10 V ± 8 mV
8 V ± 7 mV
4 V ± 5 mV
0 V Q ± 3 mV
-4V ± 5 mV
-8 V ± 7 mV
-10 V ± 8 mV
-12 V ± 9 mV
-16 V Q ± 11 mV
Output Current Quick Check Error from Nominal (90 day)
20 mA Q ± 23 μA
15 mA ± 18.5 μA
10 mA ± 14 μA
5 mA ± 9.5 μA
0 mA
[1]
5 μA
-5 mA ± 9.5 μA
-10 mA ± 14 μA
-15 mA ± 18.5 μA
-20 mA Q ± 23 μA
[1] Apply a measured “0” offset to this measurement.