EasyManua.ls Logo

Keysight Technologies 8163A/B - Page 135

Keysight Technologies 8163A/B
306 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide 135
Measurement Operations & Settings 4
command: [:SOURce[n]][:CHANnel[m]]:FREQuency:GRID
syntax: [:SOURce[n]][:CHANnel[m]]:FREQuency:GRID<wsp>
<value>[HZ|KHZ|MHZ|GHZ|THZ]|MIN|MAX|DEF
description: Set the grid spacing for the frequency grid used by the instrument.
parameters: Any value in the specified range (see appropriate User's Guide)
Also allowed:
MIN - minimum programmable value
MAX - maximum programmable value
DEF - default value
response: None
example: sour1:freq:grid 50e9
affects: 81950A
Notes Only available in "Grid
Mode"
command: [:SOURce[n]][:CHANnel[m]]:FREQuency:GRID?
syntax: [:SOURce[n]][:CHANnel[m]]:FREQuency:GRID?<wsp>[MIN|MAX|DEF]
description: Returns the grid spacing for the frequency grid used by the instrument.
parameters: Optional:
MIN - minimum programmable value
MAX - maximum programmable value
DEF - default value
response: The current grid spacing value or the minimum, maximum, or default grid spacing value in Hz.
example: sour1:freq:grid? +5.00000000E+010
affects: 81950A
Notes

Table of Contents