8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide 285
GPIB Command Compatibility List 8
Changed Parameter Syntax and Semantics
Table 13 on page -285 details commands whose parameter syntax or
semantics have changed.
Table 13 Commands with Different Parameters or Syntax
Command Change Affects
SOUR:AM:FREQ/? This command does not accept the value CW, instead
use SOUR:AM:STAT ON|OFF to switch from and to CW
mode.
The commands accepts floating point values.
8153
DISP:BRIG This command now supports integers between 1 and
100, instead of float values between 1 and 0.
8153
SENS:CORR:COLL:ZERO? This command returns the last zero state, instead of the
last remote zero state.
8153
SENS:POW:REF Accepts TOMODule and TOREF for the first parameter,
instead of accepting TOA|TOB as the HP 8153A does. The
numbers 0|1|2 cannot be used, only the strings above.
8153
SENS:POW:REF:STAT:RAT Accepts TOREF,0 or values for slot,channel, instead of
accepting TOA|TOB as the HP 8153A does. The numbers
have a different meaning.
8153
SYST:DATe SYST:DATe from HP 8167/8 is not supported, but
SYST:DATe from HP 8153 is supported.
8167/8