8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide 71
Instrument Setup and Status 3
System Communicate - The :SYST:COMMunicate sub tree.
We recommend you change network settings using GPIB or the local user
interface.
Some notes on DHCP/AutoIP/DNS
If DHCP is enabled but no DHCP server is found, the instrument tries to
use AutoIP as a fallback. This may take about 2 minutes.
Depending on the available network capabilities, the instrument tries to
tell the DNS server its host name or read the host and domain named it
has been assigned.
MAC address:
The Media Access Control (MAC) number is a unique number associated
with each network adapter.
NOTE
The instrument does not close open connections when restarting the network
interface (:SYSTem:COMMunicate:ETHernet:RESTart). This means the number of
possible connections is reduced by the number of previously open connections.
However, the instrument does make sure connections are still alive. It should
release unused open connections after about two minutes.
command: :SYSTem:COMMunicate:GPIB[:SELF]:ADDRess
syntax: :SYSTem:COMMunicate:GPIB[:SELF]:ADDRess<wsp><GPIB Address>
description: Sets the GPIB address.
parameters: The GPIB Address Values allowed 0-30
21 is often reseverved by the GPIB Controller.
response: none
example: SYST:COMM:GPIB:ADDR 20

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies 8163A/B and is the answer not in the manual?

Keysight Technologies 8163A/B Specifications

General IconGeneral
BrandKeysight Technologies
Model8163A/B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals