8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide 137
Measurement Operations & Settings 4
command: [:SOURce[n]][:CHANnel[m]]:FREQuency:OFFSet
syntax: [:SOURce[n]][:CHANnel[m]]:FREQuency:OFFset<wsp>
<value>[HZ|KHZ|MHZ|GHZ|THZ]|MIN|MAX|DEF
description: Set the frequency offset used by the instrument.
parameters: Any value in the specified range (see appropriate User's Guide)
Also allowed:
MIN - minimum programmable value
MAX - maximum programmable value
DEF - default value
response: None
example: sour1:freq:offs 0.1e9
affects: 81950A
Notes Only available in "Grid Mode"
command: [:SOURce[n]][:CHANnel[m]]:FREQuency:OFFSet?
syntax: [:SOURce[n]][:CHANnel[m]]:FREQuency:OFFSet?<wsp>[MIN|MAX|DEF]
description: Returns the frequency offset used by the instrument in Hz.
parameters: Optional:
MIN - minimum programmable value
MAX - maximum programmable value
DEF - default value
response: The current frequency offset value or the minimum, maximum, or default frequency offset value in Hz.
example: sour1:freq:offs? +1.00000000E+008
affects: 81950A