8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide 139
Measurement Operations & Settings 4
command: [:SOURce[n]][:CHANnel[m]]:FREQuency:AUTO?
syntax: [:SOURce[n]][:CHANnel[m]]:FREQuency:AUTO?
description: Returns the current operation mode.
parameters: Any value in the specified range (see appropriate User's Guide)
Also allowed:
MIN - minimum programmable value
MAX - maximum programmable value
DEF - default value
response: 0: "Grid Mode" (= Auto Mode off)
1: "Auto Mode"
example: sour1:freq:auto? 0
affects: 81950A
Notes This query is equivalent to [:SOURce[n][:CHANnel[m]]:WAVelength:AUTO?
command: [:SOURce[n]][:CHANnel[m]]:MODout
syntax: [:SOURce[n]][:CHANnel[m]]:MODout<wsp>FRQ|FRQRDY|0|1
description: Sets the modulation output mode of the BNC connector on the front panel of tunable laser modules.
parameters: FRQ or 0:
FRQRDY or 1:
modulation signal is output all the time
modulation is combined with the laser-ready signal.
In this case, the output is kept low when no optical signal is output (for example, while
the laser
is settling after a change of wavelength).
response: none
example: sour0:mod 0
affects: Tunable laser sources with BNC output connector, except 81602A, 81606A, 81607A, 81608A and 81609A.