166 8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
4 Measurement Operations & Settings
command: [:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:CYCLes
syntax: [:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:CYCLes<wsp>
<value>|MIN|MAX|DEF|0
description: Sets the number of cycles.
Cannot be set while a sweep is running.
parameters: The number of cycles is an integer value.
Also allowed are: MIN: minimum programmable value
MAX: maximum programmable value
DEF: This is not the preset (*RST) default value but is half the sum of, the minimum
programmable value and the maximum programmable value
0: cycles continuously.
response: none
example: wav:swe:cycl 3
affects: All tunable laser modules, except 81950A.
command: [:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:CYCLes?
syntax: [:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:CYCLes?
[<wsp>MIN|MAX|DEF]
description: Returns the number of cycles.
parameters: none
Also allowed are: MIN: minimum programmable value
MAX: maximum programmable value
DEF: This is not the preset (*RST) default value but is half the sum of, the minimum
programmable value and the maximum programmable value
response: The number of cycles as a signed integer value.
example: wav:swe:cycl? -> +3<END>