176 8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
4 Measurement Operations & Settings
command: [:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:STOP
syntax: [:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:STOP<wsp><stop value>
[PM|NM|UM|MM|M]
description: Sets the end point of the sweep.
Cannot be set while a sweep is running.
parameters: The wavelength at which the sweep ends as a float value in meters.
If you specify no units in your command, meters are used as the default.
response: none
example: wav:swe:stop 1550nm
affects: All tunable laser modules, except 81950A.
command: [:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:STOP?
syntax: [:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:STOP?[<wsp>MIN|MAX]
description: Returns the end point of the sweep.
parameters: optional MIN Returns the minimum start wavelength available.
This value is wavelength dependent.
MAX Returns the maximum start wavelength available.
This value is wavelength dependent.
response: The wavelength at which the sweep ends as a float value in meters.
example: wav:swe:stop? -> +1.55000000E-006<END>
affects: All tunable laser modules
command: [:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:[STATe]
syntax: [:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:[STATe]<wsp>
STOP|0|STARt|1|PAUSe|2|CONTinue|3
description: Stops, starts, pauses or continues a wavelength sweep.