178 8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
4 Measurement Operations & Settings
command: [:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:[STATe]?
syntax: [:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:[STATe]?
description: Returns the state of a sweep.
parameters: none
response: +0:
+1:
Sweep is not running
Sweep is running
example: wav:swe?
-> +0<END>
affects: All tunable laser modules, except 81950A.
command: [:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:STEP:NEXT
syntax: [:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:STEP:NEXT
description: Performs the next sweep step in stepped sweep if it is paused or in manual sweep.
parameters: none
response: none
example: wav:swe:step:next
affects: All tunable laser modules, except 81950A.
command: [:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:STEP:PREVious
syntax: [:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:STEP:PREVious
description: Performs one sweep step backwards in stepped sweep if it is paused or in
manual sweep.
parameters: none
response: none
example: wav:swe:step:prev
affects: All tunable laser modules, except 81950A.