8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide 195
Measurement Operations & Settings 4
response: 4 byte Intel floating point; the averaging time in seconds
example: OUTP1:ATIM? -> 1<END>
affects: Attenuator modules with power control.
command: :OUTPutn[:CHANnel[m]]:CORRection:COLLection:ZERO
syntax: :OUTPutn[:CHANnel[m]]:CORRection:COLLection:ZERO
description: Zeros the electrical offsets of the attenuator’s integrated powermeter.
parameters: none
response: none
example: OUTP1:CORR:COLL:ZERO
affects: Attenuator modules with power control.
command: :OUTPutn[:CHANnel[m]]:CORRection:COLLection:ZERO:ALL
syntax: :OUTPutn[:CHANnel[m]]:CORRection:COLLection:ZERO:ALL
description: Zero all available powermeter channels in the mainframe.
parameters: none
response: none
example: OUTP1:CORR:COLL:ZERO:ALL
affects: Powermeter modules; attenuator modules with power control, and return loss modules.
command: :OUTPutn[:CHANnel[m]]:CORRection:COLLection:ZERO?
syntax: :OUTPutn[:CHANnel[m]]:CORRection:COLLection:ZERO?
description: Queries the status of the last :OUTPutn[:CHANnel[m]]:CORRection:COLLection:ZERO operation.
parameters: none
response: integer 0 = OK, otherwise not OK.
example: OUTP:CORR:COLL:ZER0?
-> 0<END>
affects: Attenuator modules with power control.