EasyManua.ls Logo

Keysight Technologies 8163A/B - Page 294

Keysight Technologies 8163A/B
306 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
294 8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
9 Error Codes
Old -213 “Init ignored”
Description:
The INIT:IMM command (page 89) initiates a trigger and completes a full measurement
cycle. The continuous measurement must be DISABLED. This error code is generated if
the powermeter is still in cont. measurement mode.
Old -220 "Parameter error (StatParmError)”
Description:
The user has passed a parameter that cannot be changed in this way. The device cannot
detect one of the following more specific errors:
Old -220 -220, "Parameter error (StatParmOutOfRange)"
Description:
The user has passed a parameter that exceeds the valid range for this parameter.
Old -220 "Parameter error (StatParmIllegalVal)"
Description:
The user has passed a parameter that does not match a value in a list of possible values.
New/Old/Standard Error
Number String

Table of Contents

Related product manuals