EasyManua.ls Logo

Keysight Technologies 8163A/B - Page 294

Keysight Technologies 8163A/B
306 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
294 8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
9 Error Codes
Old -213 “Init ignored”
Description:
The INIT:IMM command (page 89) initiates a trigger and completes a full measurement
cycle. The continuous measurement must be DISABLED. This error code is generated if
the powermeter is still in cont. measurement mode.
Old -220 "Parameter error (StatParmError)”
Description:
The user has passed a parameter that cannot be changed in this way. The device cannot
detect one of the following more specific errors:
Old -220 -220, "Parameter error (StatParmOutOfRange)"
Description:
The user has passed a parameter that exceeds the valid range for this parameter.
Old -220 "Parameter error (StatParmIllegalVal)"
Description:
The user has passed a parameter that does not match a value in a list of possible values.
New/Old/Standard Error
Number String

Table of Contents

Related product manuals