EasyManua.ls Logo

Keysight Technologies 8163A/B - Page 298

Keysight Technologies 8163A/B
306 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
298 8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
9 Error Codes
Old -300 “Internal error (StatVals Lost)”
“Internal error (StatInternalError)”
Description
These are generic device-dependent errors used when the instrument cannot detect more
specific errors.
New -301 "Module doesn't support this command (StatCmdUnknown)"
Description:
The addressed module does not support the SCPI command.
Example:
When a command from the SENSe SCPI tree is sent to a fixed or tunable laser source.
New -302 "Internal timeout error (StatTimedOut)"
Description:
A command has not returned in the expected time.
New -303 "Module slot empty or slot / channel invalid"
Description:
The user has send a command to an empty slot.
New -304 "Command was aborted (StatAborted)"
Description:
The command has been interrupted by another event.
New -305 "Internal messaging error (StatCmdError)"
"Internal messaging error (StatCmdNotAllowed)"
"Internal messaging error (StatWrongLength)"
"Internal messaging error (StatWrongReceiver)"
"Internal messaging error (StatBufAllocError)"
"Internal messaging error (StatDPRamFull)"; }
"Internal messaging error (StatSemError)"
Description:
An error has occured in the instrument communication system. Please report this error
with a description of the circumstances that generated the error and the configuration of
the system.
New -306 "Channel doesn't support this command (StatCmdUnknownForSlave)"
Description:
Slave channels have limited functionality. The module supports this command, but the
command must be sent to the master channel.
New -307 "Channel without head connection (StatHeadless)"
Description:
The channel supports this command, but it cannot be executed because the optical
measurement head is not plugged into the interface module.
Standard -310 “System error”
[Indicates that some error, termed “system error” by the device, has occurred. This code
is device-dependent.]