52 8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
3 Instrument Setup and Status
command: *RST
syntax: *RST
description: The ReSeT command *RST sets the mainframe and all modules to the reset setting (standard setting) stored internally.
Pending *OPC? actions are cancelled.
The instrument is placed in the idle state awaiting a command. The *RST command clears the error queue.
The *RST command is equivalent to the *CLS command AND the syst:preset command.
The following are not changed:
GPIB (interface) state
Instrument interface address
Output queue
Service request enable register (SRE)
Standard Event Status Enable Mask (SESEM)
parameters: none
response: none
example: *RST
command: *STB?
syntax: *STB?
description: The STatus Byte query *STB? returns the contents of the Status Byte register.
parameters: none
response: The bit value for the register (a 16-bit signed integer value):
Bit Mnemonic Decimal Value
7 (MSB) Operation Status (OSB) 128
6 Not used 0
5 Event Status Bit (ESB) 32
4 Message Available (MAV) 16
3 Questionable Status (QSB) 8
2 Not used 0
1 Not used 0
0 Not used 0
example: *STB?
-> 128<END>

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies 8163A/B and is the answer not in the manual?

Keysight Technologies 8163A/B Specifications

General IconGeneral
BrandKeysight Technologies
Model8163A/B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals