EasyManua.ls Logo

Keysight Technologies 8163A/B - Page 54

Keysight Technologies 8163A/B
306 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
54 8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
3 Instrument Setup and Status
0 Selftest failed on Slot 0 1
If 16 is returned, the module in slot 4 has failed.
If 18 is returned, the modules in slots 1 and 4 have failed.
A value of zero indicates no errors.
example: *TST? -> 0<END>
command: *WAI
syntax: *WAI
description: The WAIt command prevents the instrument from executing any further commands until the current command has finished
executing. Some module firmware includes commands that set a "StatNOPC" flag during execution to indicate that the
module is busy. *WAI blocks the GPIB bus to all commands until every module hosted by the instrument is no longer busy.
All pending operations, are completed during the wait period.
parameters: none
response: none
example: *WAI

Table of Contents