8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide 59
Instrument Setup and Status 3
parameters: The bit value for the OSESM as a 16-bit signed integer value (0 .. +32767)
The default value is 0.
response: none
example: stat:oper:enab:level1 128
command: :STATus:OPERation:ENABle:LEVel1?
syntax: :STATus:OPERation:ENABle:LEVel1?
description: Returns the OSESM for the OSESR for slots 15 - 17 of the 8166A/B Lightwave Measurement System
parameters: none
response: The bit value for the operation enable mask as a 16-bit signed integer value (0 .. +32767)
example: stat:oper:enab:level1?
-> +128<END>
command: :STATusn:OPERation[:EVENt]?
syntax: :STATusn:OPERation[:EVENt]?
description: Returns the Operational Slot Status Event Register (OSSER) of slot n.
parameters: none
response: The results for the individual slot events (a 16-bit signed integer value, where 0 £ value £ 32767):
Bit
8-15
7
6
5
4
3
2
1
0
Mnemonic
Not used
Slot n: offset (λ) type bit 2
Slot n: offset (λ) type bit 1
Slot n: offset (λ) has been enabled
Slot n: shutter has been opened
Slot n: Zeroing ongoing
Not used
Slot n: Coherence Control has been switched on
Slot n: Laser has been switched on
Decimal Value
0
128
64
32
16
8
0
2
1
example: stat1:oper? -> +0<END>
command: :STATusn:OPERation:CONDition?
syntax: :STATusn:OPERation:CONDition?
description: Returns the Operational Slot Status Condition Register of slot n.