EasyManuals Logo

Keysight Technologies 8163A/B User Manual

Keysight Technologies 8163A/B
306 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #61 background imageLoading...
Page #61 background image
8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide 61
Instrument Setup and Status 3
command: :STATusn:OPERation:ENABle?
syntax: :STATusn:OPERation:ENABle?
description: Returns the OSSEM of slot n
parameters: none
response: The bit value for the OSSEM as a 16-bit signed integer value (0 .. +32767)
example: stat:oper:enab?
-> +128<END>
command: :STATus:PRESet
syntax: :STATus:PRESet
description: Presets all bits in all the enable masks for both the OPERation and QUEStionable status systems to 0, that is, OSSEM,
QSSEM, OSESM, and QSESM.
parameters: none
response: none
example: stat:pres

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies 8163A/B and is the answer not in the manual?

Keysight Technologies 8163A/B Specifications

General IconGeneral
BrandKeysight Technologies
Model8163A/B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals