2 Verification and Adjustments
74 Keysight E36200 Series Service Guide
Test Record Forms
– “Test record form - Keysight E36231A” on page 74
– “Test record form - Keysight E36232A” on page 76
– “Test record form - Keysight E36233A” on page 78
– “Test record form - Keysight E36234A” on page 80
Table 2-11 Test record form - Keysight E36231A
E36231A Report number _______________ Date _______________
Description DUT setting Lower limit Result Upper limit
Constant voltage tests
Voltage programming
Zero voltage output (V
0
)
0 V, 20 A –0.004 V ____________ 0.004 V
Maximum voltage output (V
max
)
30 V 20 A 29.987 V ____________ 30.013 V
Voltage readback
Zero voltage measured over interface 0 V, 20 A
V
0
– 0.004 V
____________
V
0
+ 0.004 V
Maximum voltage measured over interface 30 V 20 A
V
max
– 0.013 V
____________
V
max
+ 0.013 V
CV load regulation (V
load
- V
noload
)
30 V, Max A –5 mV ____________ 5 mV
10 V, Max A –3 mV ____________ 3 mV
CV line regulation (V
lowline
- V
highline
)
30 V, Max A –5 mV ____________ 5 mV
10 V, Max A –3 mV ____________ 3 mV
CV ripple and noise
peak-to-peak
30 V, Max A - ____________ 4.5 mV
10 V, Max A - ____________ 4.5 mV
rms
30 V, Max A - ____________ 0.35 mV
10 V, Max A - ____________ 0.35 mV
Transient response
30 V, Max A - ____________ 50 μs
10 V, Max A - ____________ 50 μs