Verification and Adjustments 2
Keysight E36200 Series Service Guide 85
Calibration message
You can use the CALibration:STRing command to store a message of up to 40
characters in calibration memory. For example, you could store the last calibration
date, the calibration due date, or contact information for the person responsible
for calibration. The calibration message is not affected by a power cycle or *RST.
You can only store the calibration message when the instrument is unsecured, but
you can execute the CALibration:STRing? query regardless of whether the
instrument is secured. A new calibration message overwrites the previous
message, and messages over 40 characters are truncated.
Saving calibration data
You must always save new calibration data before cycling instrument power or
leaving the calibration state with the Auto Save feature off. To save calibration
data, send CALibration:SAVE or save the calibration data from the front panel.
Calibration auto save
The instrument includes a calibration Auto Save feature. This feature
automatically saves the calibration data to non-volatile memory and increments
the calibration count when you exit the calibration state.
To enable or disable the calibration auto save feature, send CALibration:ASAVe
ON or CALibration:ASAVe OFF. To query the calibration auto save state, send
CALibration:ASAV?.