Verification and Adjustments 2
Keysight E36200 Series Service Guide 79
Constant current tests
Current programming
Zero current output (I
0
)
Max V, 0 A –0.0060 A ____________ ____________ 0.0060 A
Maximum current ouput (I
max
)
Max V, 20 A 19.974 A ____________ ____________ 20.026 A
Current readback
Zero current measured over interface Max V, 0 A
I
0
– 0.0060 A
____________ ____________
I
0
+ 0.0060 A
Maximum current measured over
interface
Max V, 20 A
I
max
– 0.026 A
____________ ____________
I
max
+ 0.026 A
CC load regulation (I
load
- I
short
)
Max V, 6.667 A –0.9167 mA ____________ ____________ 0.9167 mA
Max V, 20 A –2.250 mA 2.250 mA
CC line regulation (I
lowline
- I
highline
)
Max V, 6.667 A –0.9167 mA ____________ ____________ 0.9167 mA
Max V, 20 A –2.250 mA ____________ ____________ 2.250 mA
Low range current measurement
Maximum low range current measured
over interface
30 V, 0.1 A
I
low max
– 410 μA
____________ ____________
I
low max
+ 410 μA
Table 2-13 Test record form - Keysight E36233A (continued)
E36233A Report number _______________ Date _______________
Description DUT setting Lower limit
Result
Upper limit
Output 1 Output 2