Error Messages 4
E364xA User’s and Service Guide 155
Calibration Error Messages
The following errors indicate failures that may occur during a calibration. Refer to
the Service Guide for more information.
Table 4-3 Calibration error messages
701
Cal security disabled by jumper
The calibration security feature has been disabled with a jumper inside the
power supply. When applicable, this error will occur at power-on to warn you
that the power supply is unsecured.
702
Cal secured
The power supply is secured against calibration.
703
Invalid secure code
An invalid calibration security code was received when attempting to unsecure
or secure the power supply. You must use the same security code to unsecure
the power supply as was used to secure it, and vice versa. The security code may
contain up to11 alphanumeric characters. The first character must be a letter.
704
Secure code too long
A security code was received which contained more than 11 characters.
705
Cal aborted
A calibration in progress is aborted when you press any front-panel key, send a
device clear, or change the local/remote state of the instrument.
706
Cal value out of range
The specified calibration value (CALibration:VALue) is not valid for the present
measurement function and range.
708
Cal output disabled
Calibration is aborted by sending OUTP OFF command during calibrating a
output.