Service and Maintenance 8
E364xA User’s and Service Guide 229
14 Enter the computed value (DVM reading divided by the shunt resistance) by
using the knob and resolution selection keys.
15 Save the change and select the high current calibration point.
16 Enter the computed value (DVM reading divided by the shunt resistance) by
using the knob and resolution selection keys.
17 Save the new current calibration constants and exit the calibration mode.
If the calibration fails, an error message will be displayed for one second and
you will hear a beep, and then return to the current calibration mode again. A
CURRENT CAL message is displayed.
I MI 1.5400
I HI 2.8000
I HI 2.7890 A
CAL MODE