Installation Note N9030-90077 31
Option BBA, Analog Baseband I/Q Inputs Upgrade Kit
Verify the Installation
1. Turn on the analyzer and wait for the instrument to fully boot up.
2. Verify the spectrum analyzer application loads and sweeps as expected. Press System, Show
Hard ware on the analyzer and verify that the Main board and the BBIQ Interface Board identify
themselves as:
BBIQ
BBIQ Interface
3. Press System, Show System on the analyzer to display a list of all installed options. You
should see N9030B-BBA.
Before performing adjustments and performance verification tests using the N7814A Calibration
and Adjustment Software for Keysight RF and Microwave Instruments, press MODE/MEAS, I/Q
Analyzer (Basic), OK. Press Input/Output, Select Input, I/Q, I/Q, I/Q Cable Calibrate... on the
analyzer and follow the on-screen directions.
If the newly installed hardware is functioning properly the cable cal should work fine. Press
MODE/MEAS, Spectrum Analyzer, Swept SA, OK. Press Input/Output, Select Input, RF.

Other manuals for Keysight Technologies N9030B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N9030B and is the answer not in the manual?

Keysight Technologies N9030B Specifications

General IconGeneral
BrandKeysight Technologies
ModelN9030B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals