Security Features & Document of Volatility 26
Memory Sanitization and Removal Procedures
Instrument Sanitization Procedures
16 Make a note of the current software version, because the instrument’s software may
have been updated inside the secure area. To determine this, press System, Show
System, and write down the Instrument Software Revision. Attach this note to the
instrument.
17 Remove Disk Drive #1 and retain in controlled area.
For details of how to remove the Disk Drive, see “Disk Drive Removal Procedure” on
page 39.
18 Physically remove instrument from controlled area. Without the Disk Drive, instrument
is sanitized.
19 Install secondary non-classified Disk Drive #2, which was previously prepared in
Step 12.
Power on instrument.
20 When the instrument has booted up, press System, Show System, and confirm
that the Instrument Software Revision is the same as noted in Step 16. If not, upgrade
the instrument software to the same version as recorded in Step 16.
21 Instrument is now operational with original calibration data. Deliver to service center
for cal/repair.
22 Service center may or may not generate new calibration files on Disk Drive, depending
on whether an adjustment is performed.
This procedure assumes that new cal data was generated.
23 Instrument returned to customer’s non-controlled area.
24 Verify Instrument Software Revision, to determine whether the service center upgraded
the software. See Step 2 and Step 12.
25 Back up (archive) calibration file from Disk Drive to portable read-only media.
For details of the procedure, see “Archiving and Restoring Factory Calibration Data
Files” on page 31.
26 Remove the non-classified Disk Drive #2 and retain outside of controlled area. If
necessary, update the documented software revision on the drive label.
For details of how to remove the Disk Drive, see “Disk Drive Removal Procedure” on
page 39.
27 Physically deploy instrument back into controlled area, without a Disk Drive.
Bring read-only media containing calibration file (created in Step 25) into controlled
area.
28 Re-insert classified Disk Drive #1 into instrument, and verify Instrument Software
Revision, as in Step 2.
29 If the Instrument Software Revision does not match the revision documented in step
24 (because the software was updated at service center), then a software update must
be performed, to force the controlled-area drive to match the non-controlled-area
drive that was used to calibrate the instrument.
Step Task

Other manuals for Keysight Technologies X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies X Series and is the answer not in the manual?

Keysight Technologies X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelX Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals