Memory Sanitization and Removal Procedures
Instrument Sanitization Procedures
Replacement of Disk Drive (Instruments with PC6S or PC7S)
IMPORTANT Ensure that the instrument software revision on the secondary non-classified
disk drive matches that of the classified drive in the controlled area.
If the non–classified drive receives a software upgrade, or if the non-classified
drive is replaced with a drive that contains a newer revision of software than
that of the classified drive, the classified drive will require a software upgrade
(inside the controlled area) to match the non-classified drive.
Items Required
To perform these tasks in a controlled area, you will need:
—1x Instrument
2x Disk Drives for the instrument.
One drive (Disk Drive #1) stays permanently within the controlled area, and
the other (Disk Drive #2) stays permanently outside it
Procedure
Steps below that appear on a yellow background take place in the controlled
area. All other steps take place in a non-controlled area.
30 Restore calibration file (Step 27) from read-only media to instrument.
31 Done. Go to Step 15.
Step Task
Step Task
1 Receive new instrument from factory.
The instrument shipment may include a spare disk drive (additional drive) prepared by
the factory, which contains the same calibration data and software revision as the disk
drive in the instrument. If such a factory-prepared disk drive is present, you can skip
steps 3, 4, 6, 7, 9, and 10 below.
2 Power up the instrument and verify the Instrument Software Revision.
For instructions on how to verify the currently installed software revision, see
“Determining Instrument Software Revision” on page 9.
Write down the software revision.
3 If the instrument’s software is not the latest version, update the instrument software to
the latest version available. For instructions, see “Is your product software
up-to-date?” on page 3
Write down the software version.

Other manuals for Keysight Technologies X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies X Series and is the answer not in the manual?

Keysight Technologies X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelX Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals